Specifications |
Parameter |
Minimum |
Typical |
Maximum |
Units |
Notes |
General |
Pressure Range |
-30 |
|
300 |
mmHg |
|
Overpressure |
125 |
|
|
PSI |
|
Electrical (22°C unless otherwise stated) |
Input Excitation |
1 |
|
10 |
VDC |
Calibrated to 6 VDC |
Risk Current |
|
|
2 |
µA |
`5 |
Input Impedance |
200 |
|
|
Ω |
|
Output Impedance |
285 |
|
315 |
Ω |
|
Environmental |
Temp (Comp/Operating) |
15 |
|
40 |
°C |
|
Temperature (Storage) |
-25 |
|
70 |
°C |
|
Humidity |
10 |
|
90 |
% |
|
Light Sensitivity |
|
|
1 |
mmHg |
|
Operating Product Life |
168 |
|
|
Hours |
|
Shelf Life |
3 |
|
|
Years |
|
Mechanical |
Weight |
|
|
2 |
Grams |
|
Volume Displacement |
|
|
0.02 |
mm3 |
|
Media Interface |
|
|
|
|
Medical grade, dielectric gel |
Gel Tube Interface Material |
|
|
|
|
Polycarbonate |
Performance |
Assumes 6 VDC, 22°C and after five second warm-up unless otherwise specified. |
Offset |
-25 |
0 |
25 |
mmHg |
Zero |
Sensitivity |
4.95 |
5 |
5.05 |
µ/V/V/mmHg |
|
Calibration |
97.5 |
100 |
102.5 |
mmHg |
Output of sensor with no pressure applied and a 150 KΩ resistor shorted across + IN to + OUT |
Symmetry |
|
|
5 |
% |
|
Accuracy (-30 to 50 mmHg) |
`-1 mmHg -1% of Reading 0 |
|
+1mmHg +1% of Reading |
|
Combined effect of sensitivity, repeatability, nonlinearity and hysteresis errors. |
Accuracy (50 to 300 mmHg) |
-3 |
0 |
3 |
% output |
Combined effect of sensitivity, repeatability, nonlinearity and hysteresis errors. |
Temp Coeff - Zero |
-0.3 |
0 |
0.3 |
mmHg/°C |
Over a temperature range of 15°C to 40°C. |
Temp Coeff - Sensitivity |
-0.1 |
0 |
0.1 |
%/°C |
Over a temperature range of 15°C to 40°C. |
Frequency Response |
1,200 |
|
|
Hz |
Previously qualified, not tested in production |
Phase Shift |
|
|
5 |
degrees |
Previously qualified, not tested in production |
Offset Drift |
|
|
1 |
mmHg/8 hrs |
Normalized offset/bridge voltage - 8 hours after 20 second warm-up, Previously qualified, not tested in production |
Standard Part Numbers : 901051001