Piezoresistive Pressure Sensors

Piezoresistive Pressure Sensors ( 압저항방식 압력센서, Blood Pressure )

DaiL IBC 2010. 3. 18. 16:01

Blood Pressure Packages

BP Series disposable medical,compensated

 

The BP Series is ideal for high-volume, disposable
medical applications and is compliant with the AAMI
BP22 specification for blood pressure monitoring.

 

 

 

 

         www.dailibc.com

 

FEATURES

Range

-30 to 300 mmHg (-.58 to 5.8 pso; 40 to 400 mbar;
4 to 40 KPa). Also available in higher pressures.

Type

Gage

Media

Air, gasses and liquids

Packaging

Die and packaged die; with and without gel; singulated
and snapstrate form

Flexibility

Die only: sensitivity, resistance, bridge, constraing, etc.


BENEFITS

Performance

Enjoy best-in-class performance due to Merit's
proprietary Sentium technology

Cost

Save money over time with high-performing die

Security

Feel confident doing business with an experienced
company backed by a solid parent company (NASDAQ: MMSI)

Speed

Get to market quickly with creative and flexible solutions.

Service

Experience prompt, personal and professional support.

 

 

 

 

 

SPECIFICATIONS
 

Parameter

Minimum

Typical

Maximum

Units

Notes

General

 

Pressure Range

-30

 

300

mmHg

 

 

Overpressure

125

 

 

PSI

 

Electrical (22°C unless otherwise stated)

 

Input Excitation

1

 

10

VDC

Calibrated to 6 VDC

 

Risk Current

 

 

2

µA

5

 

Input Impedance

200

 

 

Ω

 

 

Output Impedance

285

 

315

Ω

 

Environmental

 

Temp (Comp/Operating)

15

 

40

°C

 

 

Temperature (Storage)

-25

 

70

°C

 

 

Humidity

10

 

90

%

 

 

Light Sensitivity

 

 

1

mmHg

 

 

Operating Product Life

168

 

 

Hours

 

 

Shelf Life

3

 

 

Years

 

Mechanical

 

Weight

 

 

2

Grams

 

 

Volume Displacement

 

 

.02

mm3

 

 

Media Intercade

 

 

 

 

Medial grade, dielectric gel

 

Gel Tube Interface Material

 

 

 

 

Polycarbonate

Performance 1

 

Offset

-25

0

25

mmHg

Zero

 

Sensitivity

4.95

5

5.05

µ/V/V/mmHg

 

 

Calibration

97.5

100

102.5

mmHg

2

 

Symmetry

 

 

5

%

 

 

Accuracy (-30 to 50 mmHg)

-1 mmHg -1%

0

+1mmHg +1%

 

6

 

Accuracy (50 to 300 mmHg)

-3

0

3

% output

6

 

Temp Coeff - Zero

-3

0

0.3

mmHg/°C

3

 

Temp Coeff - Sensitivity

-1

0

0.3

mmHg/°C

3

 

Frequency Response

1,200

 

 

Hz

5

 

Phase Shift

 

 

5

degrees

5

 

Offset Drift

 

 

1

mmHg/8 hrs

4, 5



1. Assumes 6 VDC, 22°C and after five second warm-up unless otherwise specified. 3. Over a temperature range of 15°C to 40°C. 5. Previously qualified, not tested in production
2. Output of sensor with no pressure applied and a 150 KΩ resistor shorted across + IN to + OUT 4. Normalized offset/bridge voltage - 8 hours after 20 second warm-up 6. Combined effect of sensitivity,repeatability, nonlinearity and hysteresis errors.

 

         www.dailibc.com